• NIBP monitor
• Adult/Pediatric
• Non-invasive blood pressure: systolic, mean, and diastolic pressures at preset levels
• Alarm and visual signals when pressure goes outside preset limits
• Pulse rate
Product Specifications
• H x W x D - 22.4 x 16.9 x 16.9 cm / 8.75 x 6.625 x 6.625 in
• Weight - 4.3 kg / 9.5 lbs
• Determination Time - 20 to 45 seconds typical; 120 seconds maximum
• Prior Data - Monitor stores 99 minutes of data (up to 100 readings within this period) while monitor is on.
• Operating Modes - AUTOMATIC and MANUAL Modes. Pressing CANCEL will stop determination in any mode.
• Power Requirements - Input power: 0.4 amps max at 100, 120 VAC amps max. at 220, 240 VAC
• Input voltage - Domestic: 120 VAC/60 Hz, 104-132 VAC, 47-63 Hz. International: 100 VAC/50 Hz (nom.), 88-112 VAC, 47-63 Hz or 230 VAC / 50 Hz (nom.), 195-264 VAC, 47-63 Hz.
• Battery - Six hour minimum operation (5 minutes auto cycle with adult cuff at 25°C) with full charge. Battery is 90% to 100% charged after 16 hours of charging.
• Accessory Connection - ear and panel data interface connector provides serial data communications interface at 600 BAUD.
For service, parts, PM or rental see our Services section